Underlying Rhythm Questions : What is happening in this EGM? Answer: This is an underlying rhythm freeze for an Abbott dual chamber pacemaker. What is the underlying rhythm? Mobitz II block. Notice the sinus(atrial) rate is approximately 90bpm and the ventricular rate is approximately half that at 48bpm.
Sorin Device Troubleshooting Questions : The following EGMs are from a routine Sorin dual chamber pacemaker interrogation. Both EGMs were stored as “V Bursts”. What is your analysis? The patient has complete heart block. Sensing, threshold, and impedance measurements were all stable. AP 3%, VP >99% Answer: The first recorded episode contains NSVT with appropriate… Continue reading Sorin Device Troubleshooting
PMT Questions : What is happening in this EGM? Answer: This is an example of a PMT episode that is successfully terminated with the algorithm AP on PMT (Abbott). The PMT detection rate is 120bpm. The device detected eight consecutive VP-AS intervals with a rate higher than the PMT detection rate (appx. 120), the device… Continue reading PMT
Atrial Lead Diagnostic Questions : What is happening in this EGM? This patient presented to device clinic for a routine follow up interrogation. The patient’s AP% was 1.5%, BP 89%, and PVC count 8.9%. The patient stated “periodically feeling “jolts” that are uncomfortable. The following EGM is the presenting rhythm freeze. Answer: Some atrial under… Continue reading Atrial Lead Diagnostic